Summary

Product identification

Testomat 2000® CLT - Process-photometer for monitoring total chlorine (CLT).

Application

  • Monitoring of chlorination systems for drinking water/swimming pool water
  • Protection for reverse osmosis membranes
  • Monitoring of biocides and conditioning agents containing chlorine

Operating limits

  • Ambient temperature: 10–45 °C (50 - 104°F)
  • Water temperature: 10 to 40 °C (50 to 104°F)

Measuring range

  • Total chlorine: 0.00 - 0.99 mg/l and 1.0 - 2.5 mg/l
  • Or free chlorine: 0.00 - 0.99 mg/l and 1.0 - 2.5 mg/l

Feed water

  • Operating pressure:
    • 1 to 8 bar (15 to 116 psi) or
    • 0.3 to 1 bar (5 to 15 psi), after removing of the valve body out of the regulator/filter housing
  • Water inlet hose: 6 mm external diameter / 4 mm internal diameter

Product water

  • Water outlet hose: 6 mm external diameter / 4 mm internal diameter

Electrical

  • Power supply: 230–240 VAC, 115 VAC, 24 VAC (50–60 Hz)
  • Power consumption: max. 30 VA
  • Protection class: I
  • Degree of protection: IP 65

Mechanical / hydraulic

  • Dimensions (W x H x D): 380 x 480 x 280 mm
  • Weight: approx. 9.5 kg
  • Menu languages: German, English, French, Italian (others upon request)
  • Option: booster pump (see accessories)