
Datasheet summary
Testomat 2000 CLT self clean
Engineering summary from PDF text extraction for Testomat 2000 CLT self clean. Verify every value with the OEM datasheet.
Summary
Product identification
The process photometer Testomat 2000® CLT self clean is an enhanced version of the Testomat 2000® CLT, featuring an additional pump for detergent dosing. This addition allows for the cleaning of the measuring chamber and outlet hose.
Application
- Disinfectant monitoring for drinking water and process water
- Medical technology (dialysis)
Measuring range
- Total chlorine: 0.00 - 0.99 mg/l
- Total chlorine: 1.0 - 2.5 mg/l
Technical Data
Electrical
- Power supply: 230–240 VAC, 115 VAC, 24 VAC (all 50–60Hz)
- Power consumption: max. 30 VA
- Protection class: I
Mechanical / hydraulic
- Degree of protection: IP 65
- Dimensions (W x H x D): 380 x 480 x 280 mm
- Weight: approx. 9.5 kg
- Option: booster pump (see accessories)
Operating limits
- Ambient temperature: 10–45 °C (50 - 104°F)
- Mains water supply operating pressure: 1 to 8 bar (15 to 116 psi) or 0.3 to 1 bar (5 to 15 psi) after removing the valve body from the regulator/filter housing.
- Water temperature: 10 to 40 °C (50 to 104°F)
Feed water
- Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter
- Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter
General
- Menu languages: German, English, French (others upon request)
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.