Datasheet summary
Testomat 2000 DUO
Engineering summary from PDF text extraction for Testomat 2000 DUO. Verify every value with the OEM datasheet.
Summary
Product identification
The Testomat 2000® DUO is a special design of the Testomat 2000®. It can measure two different parameters simultaneously, such as two measuring ranges for water hardness, or water hardness and carbonate hardness. Monitoring of two measuring points is possible with automatic measuring point switchover and possible suppression of the input contact.
Application
- Use in two circuits with different hardnesses
- Measurement of inlet and outlet hardness
Operating limits
Feed water
- Operating pressure: 1 to 8 bar (15 to 116 psi) or 0.3 to 1 bar (5 to 15 psi) after removing the valve body from the regulator/filter housing
- Water inlet hose: 6 mm external diameter / 4 mm internal diameter
- Water outlet hose: 6 mm external diameter / 4 mm internal diameter
- Water temperature: 10 to 40 °C (50 to 104°F)
Product water
- Measuring range (depending on the choice of indicator):
- 0.05–25 °dH Water hardness
- 0.5–20 °dH Carbonate hardness
- 1–15 mmol/l p-value
- 0.05–0.5 mmol/l minus m-value
Electrical
- Power supply: 230–240 VAC, 115 VAC, 24 VAC (all 50–60 Hz)
- Power consumption: max. 30 VA
- Protection class: I
- Degree of protection: IP 65
Mechanical / hydraulic
- Dimensions (W x H x D): 380 x 480 x 280 mm
- Weight: approx. 9.5 kg
- Ambient temperature: 10–45 °C (50 - 104°F)
- Option: booster pump (see accessories)
- Menu languages: German, English, French, Italian, Polish (others upon request)
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.