Testomat 2000 Fe

Engineering summary from PDF text extraction for Testomat 2000 Fe. Verify every value with the OEM datasheet.

Summary

Product identification

The Testomat 2000® Fe is a wet-chemical online monitor for the photometric analysis of iron content.

Operating limits

  • Ambient temperature: 10–45 °C (50 - 104°F)
  • Water temperature: 10 to 40 °C (50 to 104°F)

Feed water

  • Operating pressure: 1 to 8 bar (15 to 116 psi) or 0.3 to 1 bar (5 to 15 psi) (after removing the valve body from the regulator/filter housing)
  • Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter
  • Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter

Product water

Not specified.

Electrical

  • Power supply: 230–240 VAC, 115 VAC, 24 VAC (all 50–60Hz)
  • Power consumption: max. 30 VA
  • Protection class: I
  • Degree of protection: IP 65

Mechanical / hydraulic

  • Dimensions (W x H x D): 380 x 480 x 280 mm
  • Weight: approx. 9.5 kg
  • Menu languages: German, English, French, Italian, Polish (others upon request)
  • Option: Booster pump (see accessories)

Measurement

  • Parameter: Iron-II and Iron-III
  • Measuring range:
    • 0.00 - 0.65 mg/l (resolution of 0.01)
    • 0.7 - 1.0 mg/l (resolution of 0.1)

Disclaimer: This summary is based on the provided text extraction. It may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.