
Datasheet summary
Testomat 2000
Engineering summary from PDF text extraction for Testomat 2000. Verify every value with the OEM datasheet.
Summary
Product identification
The Testomat 2000® is an automatic titrator for determining water hardness, carbonate hardness, the p-value, or the minus m-value.
Application
- Water treatment plants
- Water blending plants
- Drinking water plants
- Water softening plants
- Decarbonization plants
- Desalination plants
- Boiler houses
- Cooling towers
Operating limits
Measuring range
- Water hardness: 0.05-25 °dH
- Carbonate hardness: 0.5-20 °dH
- p-value: 1-15 mmol/l
- Minus m-value: 0.05-0.5 mmol/l (depending on the choice of the indicator)
Feed water
- Operating pressure: 1 to 8 bar (15 to 116 psi) or 0.3 to 1 bar (5 to 15 psi) after removing the valve body from the regulator/filter housing.
- Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter.
- Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter.
- Water temperature: 10 to 40 °C (50 to 104 °F).
Electrical
- Power supply: 230–240 VAC, 115 VAC, 24 VAC, all 50–60 Hz.
- Power consumption: max. 30 VA.
- Protection class: I.
- Degree of protection: IP 65.
Mechanical / hydraulic
- Dimensions (W x H x D): 380 x 480 x 280 mm.
- Weight: approx. 9.5 kg.
- Ambient temperature: 10–45 °C (50 - 104°F).
- Option: Booster pump (see accessories).
- Menu languages: German, English, French, Italian, Polish, Dutch (others upon request).
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.