Testomat 2000

Engineering summary from PDF text extraction for Testomat 2000. Verify every value with the OEM datasheet.

Summary

Product identification

The Testomat 2000® is an automatic titrator for determining water hardness, carbonate hardness, the p-value, or the minus m-value.

Application

  • Water treatment plants
  • Water blending plants
  • Drinking water plants
  • Water softening plants
  • Decarbonization plants
  • Desalination plants
  • Boiler houses
  • Cooling towers

Operating limits

Measuring range

  • Water hardness: 0.05-25 °dH
  • Carbonate hardness: 0.5-20 °dH
  • p-value: 1-15 mmol/l
  • Minus m-value: 0.05-0.5 mmol/l (depending on the choice of the indicator)

Feed water

  • Operating pressure: 1 to 8 bar (15 to 116 psi) or 0.3 to 1 bar (5 to 15 psi) after removing the valve body from the regulator/filter housing.
  • Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter.
  • Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter.
  • Water temperature: 10 to 40 °C (50 to 104 °F).

Electrical

  • Power supply: 230–240 VAC, 115 VAC, 24 VAC, all 50–60 Hz.
  • Power consumption: max. 30 VA.
  • Protection class: I.
  • Degree of protection: IP 65.

Mechanical / hydraulic

  • Dimensions (W x H x D): 380 x 480 x 280 mm.
  • Weight: approx. 9.5 kg.
  • Ambient temperature: 10–45 °C (50 - 104°F).
  • Option: Booster pump (see accessories).
  • Menu languages: German, English, French, Italian, Polish, Dutch (others upon request).