Summary
Product identification
The Testomat® 808 is a limit value monitoring device for water hardness, serving as a further development of the Testomat® F-BOB.
Application
- Reverse osmosis plants
- Soft water for commercial purposes
- Pure water production plants
- Galvanization
Operating limits
Measuring range
- Water hardness: 0.02 – 3 °dH (0.4….54 ppm CaCO₃)
Ambient temperature
- 10 – 40 °C; 50 - 104°F
Water temperature
- 10 to 40 °C; 50 to 104°F
Electrical
- Power supply: Depending on product configuration: 24V / 115V / 230V, 50 – 60 Hz
- Instrument protection:
- 230 – 240 V: T0.1 A
- 115 V: T0.2 A
- 24 V: T0.8 A
- Mains protection for consumers: max. 4 A (n, l)
- Power consumption: max. 16 VA, without external load
- Protection class: I
- Degree of protection: IP 54
- Conformity: EN 61000-6-2, EN 61000-6-4, EN 61010-1
- Current interface: Output of defined values (5, 8, 11, 14, 17, 20 mA) for displaying status and error messages, max. load 500 Ohms
- Contact load relay: 230V / 4A AC ohm resistive load
Mechanical / hydraulic
- Dimensions: B x H x T = 364 x 314 x 138 mm
- Weight: 4350 g
Mains water supply
- Operating pressure: Depending on product configuration:
- 0.3 – 1 bar / 0.3 x 10⁵ to 1 x 10⁵ Pa
- 1 - 4 bar / 1 x 10⁵ to 4 x 10⁵ Pa
- (A pressure reducer (special accessory) should be used from 4 to 8 bar range)
- Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter
- Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter