M200 1-2 Channel Multiparameter Transmitter
Engineering summary from PDF text extraction for M200 1-2 Channel Multiparameter Transmitter. Verify every value with the OEM datasheet.
Summary
Product Identification
The Mettler Toledo M200 series are single and dual-channel multiparameter analytical transmitters for basic water and process applications. They are capable of measuring conductivity/resistivity, pH/ORP, dissolved oxygen, dissolved ozone, and flowrate. The M200 is compatible with easySense and ISM sensors.
Features
- Large backlit 4-line display
- Compact ¼ DIN panel-mounting with IP65 panel seal
- Robust ½ DIN wall-mounting with IP65 back cover
- User-selectable dual-level password protection
- USB port for configuration and data acquisition
- Plug and measure sensor inputs for all parameters
- Compatible with easySense and a range of ISM sensors
- Plug-in terminal connectors for easy wiring
- Universal AC/DC power supply
- Multiple languages supported (English, French, German, Italian, Spanish, Portuguese, Russian, Japanese)
- CE compliant, UL Listed for US and Canadian standards
Applications
- Pure and ultrapure water treatment (semiconductor rinsing, critical power/steam makeup, pharmaceutical waters)
- Semiconductor processing (rinsers, wet benches)
- Power plant cycle chemistry and stator cooling monitoring
- Pharmaceutical water monitoring (meeting USP, EP, ChP, IP, JP conductivity requirements)
- Reclaim, recycle, and wastewater treatment for contaminant detection, diversion, and neutralization
Physical and Electrical Specifications
Enclosure/Power
- Operator Interface: 4-line backlit LCD; 5 tactile keys
- ¼ DIN Model Dimensions (H x W x D): 4.01 x 4.01 x 5.51 in (102 x 102 x 140 mm)
- Max. Depth (panel mounted): 4.96 in (126 mm), excluding plug-in connectors
- ½ DIN Model Dimensions (H x W x D): 5.90 x 5.90 x 4.57 in (150 x 150 x 116 mm)
- Max. Depth (panel mounted): 3.46 in (88 mm)
- Material: Polycarbonate
- Weight (¼ DIN): 1.5 lb (0.7 kg)
- Weight (½ DIN): 2 lb (1 kg)
- UL Electrical Environment: Installation (overvoltage) Category II
- Ratings/Approvals: UL listed, CE compliant; IP 65 (½ DIN wall mount and ¼ DIN panel seal)
- EMC: Compliant with EN61326-1:2013 (Industrial Environment Emission: Class B, Immunity: Class A)
- Power: Universal 100-240 VAC, 50-60 Hz or 20-30 VDC; 5 W. Settings retained in non-volatile memory on power loss.
Environmental
- Storage Temperature: -40 to 70 °C (-40 to 158 °F)
- Ambient Temperature Operating Range: -10 to 50 °C (14 to 122 °F)
- Relative Humidity: 0 to 95%
Outputs/Inputs
- Analog Outputs: Powered 0/4-20 mA, 22 mA alarm, 500 ohms maximum load.
- Accuracy: ± 0.05 mA
- Scaling: Linear
- Relays: All contacts are potential free, with adjustable hysteresis and time delay.
- Type: SPDT
- Rating: 250 VAC/30 VDC, 3 A, resistive
- Digital Communications: USB Type B connector for remote configuration and data acquisition with TCT software.
- Discrete Input: Accepts dry contact closure for remote PID control auto/manual selection.
Channel Configuration
| Feature | One Channel | Two Channel |
|---|---|---|
| Setpoints/Alarms | 4 | 6 |
| Relays | 2 SPDT | 2 SPDT |
| Analog Output Signals | 2 | 4 |
| Discrete Inputs | 1 | 2 |
(Sensor/measurement specifications require referring to Mettler-Toledo Process Analytics easySense and Analytical Sensor datasheets.)
M200 Flow Models
- Available in one and four-channel models.
- Available in ¼ DIN enclosure only.
- Accepts input from most pulse flow sensors.
- Computes total flow and reverse osmosis % recovery (four-channel model).
- Configurable for batch control.
Measurement Specifications (Flow)
- Flowrate Range: 0 to 9999 GPM, L/min, m³/hr
- Total Flow Range: 0 to 999,999,999 Gallons, Liters, or m³
- RO % Recovery Range: 0 to 100%
- Flow Velocity Range: Equivalent ft/s, m/s
- Frequency Range: 1 to 4000 Hz
- Calculated Parameters (4-channel): Ratio, sum, and difference of two flowrates.
- Resolution: 4 significant digits, auto-ranged; up to 8 digits for total flow.
- Update Rate: Display and outputs, once per 2 seconds.
- Input Pulses: Low <1.0 volt; High >1.4 volts (36 volts max).
- Accuracy: ± 0.5 Hz
- Repeatability: ± 0.2 Hz
Channel Configuration (Flow Models)
| Feature | Single Channel | Four Channel |
|---|---|---|
| Setpoints/Alarms | 4 | 8 |
| Relays | 2 SPDT, 1 SPST NO, 1 SPST NC | 2 SPDT, 1 SPST NO, 1 SPST NC |
| Analog Output Signals | 2 | 4 |
| Discrete Inputs (totalizer reset) | 1 | 2 |
Ordering Information
M200 Multi-parameter Transmitters
| Description | Part Number |
|---|---|
| 1/4 DIN, 1-channel | 52 121 554 |
| 1/4 DIN, 2-channel | 52 121 556 |
| 1/2 DIN, 1-channel | 52 121 555 |
| 1/2 DIN, 2-channel | 52 121 557 |
(Panel and pipe mounting for 1/2 DIN models require separately ordered kits.)
M200 Flow Transmitters
| Description | Part Number |
|---|---|
| 1/4 DIN, 1-channel Flow | 30 280 748 |
| 1/4 DIN, 2-channel Flow | 30 280 749 |
Accessories
| Description | Part Number |
|---|---|
| Panel Mount Kit for 1/2 DIN models | 52 500 213 |
| Pipe Mount Kit for 1/2 DIN models | 52 500 212 |
| ISM Conductivity Calibration Module | 58 082 305 |
| Adapter Panel - M200 to 200/2000 cutout | 58 083 305 |
ISM Sensor Cables (Thornton Conductivity)
| Length | Part Number |
|---|---|
| 1 ft (0.3 m) | 58 080 270 |
| 5 ft (1.5 m) | 58 080 271 |
| 10 ft (3.0 m) | 58 080 272 |
| 15 ft (4.5 m) | 58 080 273 |
| 25 ft (7.6 m) | 58 080 274 |
| 50 ft (15.2 m) | 58 080 275 |
| 100 ft (30.5 m) | 58 080 276 |
| 150 ft (45.7 m) | 58 080 277 |
| 200 ft (61.0 m) | 58 080 278 |
| 300 ft (91.4 m) | 58 080 279 |
ISM Sensor Cables (pH/ORP/DO)
| Length | Part Number |
|---|---|
| 3 ft (1 m) | 59 902 167 |
| 10 ft (3 m) | 59 902 193 |
| 16 ft (5 m) | 59 902 213 |
| 33 ft (10 m) | 59 902 230 |
(Note: easySense sensors include cable from sensor to transmitter.)