M300 ISM 1-2 Channel Multiparameter Transmitter

Engineering summary from PDF text extraction for M300 ISM 1-2 Channel Multiparameter Transmitter. Verify every value with the OEM datasheet.

Summary

Product identification

M300 ISM Multiparameter Transmitters for Innovative Liquid Analysis. Used with UniCond™ conductivity sensors with ISM.

Applications

  • Pure and ultrapure water treatment for semiconductor rinsing, critical power/steam makeup and pharmaceutical waters.
  • Semiconductor processing in rinsers and wet benches with precise resistivity measurement and temperature compensation.
  • Power plant cycle chemistry and stator cooling monitoring with accurate temperature compensation for specific and cation conductivity, and pH.
  • Pharmaceutical water monitoring to meet USP, EP, and JP conductivity requirements.
  • Reclaim, recycle and wastewater treatment for contaminant detection, diversion, and neutralization.

Features

  • Compact 1/4 DIN panel mounting with NEMA 4X, IP65 panel seal.
  • Robust 1/2 DIN wall mounting with NEMA 4X, IP65 back cover.
  • ISM Sensors are immediately recognized at installation.
  • Measurements are available within seconds of connection.
  • ISM capabilities communicate type, model, serial number, full calibration data plus historical data.
  • On-line sensor diagnostics assure process surveillance.
  • Historical calibration and exposure information improves preventive maintenance programs.
  • Multiple languages: English, French, German, Italian, Spanish, Portuguese, Russian, Japanese.
  • CE compliant, UL Listed for US and Canadian standards.
  • "Plug and Measure" capabilities virtually eliminate operational errors.

Functional Specifications

Conductivity/Resistivity Ranges

  • 0.01 Constant Sensor: 0.002 to 500 µS/cm (2000 ohm-cm to 500 Mohm-cm)
  • 0.1 Constant Sensor: 0.02 to 50,000 µS/cm (20 ohm-cm to 50 Mohm-cm)
  • Readout in equivalent S/m ranges is selectable.
  • TDS Ranges (CaCO₃ and NaCl): Cover equivalent conductivity ranges.
  • RO % Rejection: 0-100% (computed from two conductivity measurements).

Conductivity Resolution

  • Four significant digits, auto-ranged.

pH, ORP Ranges

  • -1.00 to 15.00 pH, -1500 to 1500 mV.
  • Resolution: 0.01 pH, 1 mV.

Dissolved Oxygen Ranges

  • 0-20,000 ppb; 0-20 ppm, 0-200% saturation.
  • Resolution: 0.1 ppb.

Temperature Measurement Range

  • -40 to 200 °C (-40 to 392 °F).
  • Resolution: 0.1 °C.

Temperature Sensor Input

  • Digital from sensor.

Update rate

  • Display and outputs, once per second.

Temperature Compensation

  • Conductivity/Resistivity: Selectable as: Std (standard high purity Thornton/Light), Light 84, Std referenced to 75°C, linear %/°C, 50% glycol, 100% glycol, cation, ammonia, isopropyl alcohol, none.
  • pH: Automatic / manual for electrode output, plus adjustable solution temperature coefficient for solution ionization effects.
  • Dissolved Oxygen: Automatic, for membrane permeability and oxygen solubility.

Calibration

  • Conductivity/Resistivity: 1- or 2-point.
  • pH: 1- or 2-point, with auto buffer recognition.
  • Dissolved Oxygen: Air/comparison, zero.

pH Diagnostics

  • Selectable continuous checking of membrane resistance and reference diaphragm/junction resistance (with solution ground sensors).

Sensor maximum distance

  • 91 m (300 ft) conductivity.
  • 20 m (66 ft) pH, ORP, DO.

Performance

  • System Accuracy: Identical to sensor for Conductivity/Resistivity, pH, ORP, and Dissolved Oxygen.

Outputs

  • Channels: One or Two.
  • Setpoints/Alarms: 4 (One Channel) or 6 (Two Channel) - high, low, outside, between, USP or EP.
  • Relays: 2 (One Channel) or 2 SPDT, 1 SPST NO, 1 SPST NC, 2 SPST reed (Two Channel) - SPDT, SPST reed types.
  • Analog Output Signals: 2 (One Channel) or 4 (Two Channel).
  • Discrete Inputs: 1 (One Channel) or 2 (Two Channel).

Mechanical / Hydraulic

Enclosure/Power

  • Operator interface: 4 line back-lit LCD; 5-tactile keys.

Dimensions

  • 1/4 DIN Model (H x W x D): 4.01 x 4.01 x 5.51 in (102 x 102 x 140 mm).
    • Max. Depth, panel mounted: 4.96 in (126 mm), excluding plug-in connectors.
  • 1/2 DIN Model (H x W x D): 5.90 x 5.90 x 4.57 in (150 x 150 x 116 mm).
    • Max. Depth, panel mounted: 3.46 in (88 mm).

Material

  • Polycarbonate.

Weight

  • 1/4 DIN Models: 1.5 lb (0.7 kg).
  • 1/2 DIN Models: 2 lb (1 kg).

Environmental

  • Storage temperature: -40 to 70 °C (-40 to 158 °F).
  • Ambient temperature operating range: -10 to 50 °C (14 to 122 °F).
  • Relative humidity: 0 to 95%.

Physical and Electrical Specifications

  • Analog outputs: Powered 0/4-20 mA, 22 mA alarm, 500 ohms maximum load.
    • Accuracy: ± 0.05 mA.
    • Scaling: Linear, bi-linear, logarithmic (1,2,3 or 4 decades), auto-ranging.
  • Relays: All contacts are potential free, with adjustable hysteresis and time delay.
    • SPDT, SPST NO, SPST NC: 250 VAC/30 VDC, 3 A, resistive.
    • SPST reed: 300 VDC, 0.5 A, 10 W.
  • Digital communications: USB, type B connector, for remote configuration and data acquisition.
  • Discrete input: Accepts dry contact closure for remote PID control auto/manual selection.
  • PID Control: Auto/manual status and % output on bottom line of display. Settings include setpoint, deadband, non-linear corner points, control limits, proportional gain, integral reset time (min.), derivative rate time (min.).
  • Manual Station: Controlled by up/down arrow keys in manual mode; remote auto/manual selection by discrete input.
  • Control Output Types: One or two analog signals, relays—pulse frequency, or relays—pulse length.

Electrical

  • Power: Universal 100-240 VAC, 50-60 Hz or 20-30 VDC; 5 W.
  • Non-volatile memory: All settings are retained on power loss without batteries.
  • UL Electrical Environment: Installation (overvoltage) Category II.

Ratings/Approvals

  • UL listed, CE compliant.
  • NEMA 4X, IP 65 - 1/2 DIN wall mount and 1/4 DIN panel seal.
  • EMC Emissions: According to EN55011 Class A.

Disclaimer