Datasheet summary
DULCOTEST CDE
Engineering summary from PDF text extraction for DULCOTEST CDE. Verify every value with the OEM datasheet.
Summary
Product Identification
DULCOTEST® Sensors for Chlorine Dioxide
Product Ranges
- CDE: For clear water, e.g., disinfection of potable water.
- CDP: For rapid control of chlorine dioxide content, e.g., in bottle-washing systems, with surfactants and rapid temperature changes.
- CDR: For polluted waters, e.g., raw water for potable water generation, cooling water, and wastewater.
CDE 2-mA Chlorine Dioxide Sensor
Standard sensor for measurement of chlorine dioxide without cross-sensitivity to free chlorine. Designed for controllers with 4-20 mA input.
CDP 1-mA Chlorine Dioxide Sensor
Sensor for fast response time chlorine dioxide measurement, suitable for bottle-washing systems. Designed for controllers with 4-20 mA input. Requires external temperature compensation.
CDR 1-mA Chlorine Dioxide Sensor
Sensor for chlorine dioxide measurement in all water types, including hot and contaminated water, without cross-sensitivity to free chlorine. Designed for controllers with 4-20 mA input.
CDR 1-CAN Chlorine Dioxide Sensor
Sensor for chlorine dioxide measurement in all water types, including hot and contaminated water, without cross-sensitivity to free chlorine. Designed for connection to a CAN interface.
Operating Limits
CDE 2-mA
- Temperature: 5 … 45 °C
- Max. pressure: 1.0 bar
- Intake flow: 30…60 l/h (in DGM or DLG III)
- pH range: 4.0 … 11.0
CDP 1-mA
- Temperature: 10 … 45 °C
- Max. pressure: 3.0 bar
- Intake flow: 30…60 l/h
- pH range: 5.5 … 10.5
CDR 1-mA
- Temperature: 1 … 55 °C (short-term 60 °C)
- Max. pressure: 3.0 bar (at 30°C, in the DGMa)
- Intake flow: 30…60 l/h (in DGM or DLG III)
- pH range: 1.0 … 10.0
CDR 1-CAN
- Temperature: 5 … 45 °C
- Max. pressure: 1.0 bar
- Intake flow: 30…60 l/h (in DGM or DLG III)
- pH range: 1.0 … 10.0
Feed Water
CDE 2-mA
- Resistance to: Salts, acids, alkalis. Not surfactants.
- Typical applications: Uncontaminated drinking water (surfactant-free).
CDP 1-mA
- Resistance to: Surfactants, slight films of dirt.
- Typical applications: Process water containing surfactants (bottle washing machines).
CDR 1-mA & CDR 1-CAN
- Resistance to: Surfactants, slight films of dirt, water-soluble chemicals, solids/dirt, biofilms.
- Typical applications: Contaminated industrial/process water, surfactant-containing, cooling water, irrigation water, slightly contaminated wastewater, warm water.
Product Water
Measured Variable
- Chlorine dioxide (ClO₂)
Selectivity
- CDE 2-mA: Chlorine dioxide selective towards free chlorine, chlorite and chlorate.
- CDP 1-mA: Chlorine dioxide as against chlorite and chlorate.
- CDR 1-mA & CDR 1-CAN: Chlorite.
Cross Sensitivity
- CDE 2-mA, CDR 1-mA & CDR 1-CAN: Ozone.
- CDP 1-mA: Ozone, chlorine.
Measuring Principle
- Amperometric, 2 electrodes, membrane-covered.
Electrical
Supply Voltage
- CDE 2-mA, CDP 1-mA, CDR 1-mA: 16...24 V DC (two-wire technology)
- CDR 1-CAN: Via CAN interface (11-30 V)
Output Signal
- CDE 2-mA, CDP 1-mA, CDR 1-mA: 4…20 mA (uncalibrated, not electrically isolated)
- CDR 1-CAN: Uncalibrated, electrically isolated.
CDE 2-mA Specifics
- Output signal is temperature-compensated.
- Response time sensor t90: 120 s
CDP 1-mA Specifics
- Output signal is not temperature-compensated. Requires separate temperature measurement for compensation.
- Response time sensor t90: 60 s
CDR 1-mA Specifics
- Output signal is temperature-compensated.
- Response time sensor t90: 3 min.
CDR 1-CAN Specifics
- Output signal is temperature-compensated.
- Response time sensor t90: 3 min.
Mechanical / Hydraulic
Sensor Fitting
- DGM, DLG III (DLG II for CDP 1-mA). A mounting kit (order no. 815079) is required for initial fitting in the DLG III in-line probe housing.
Installation
- Bypass: open sample water outlet.
Measuring Ranges (mg/l)
- CDE 2-mA:
- 0.01…0.5 ppm
- 0.02…2.0 ppm
- 0.10…10.0 ppm
- CDP 1-mA:
- 0.02…2.0 ppm
- CDR 1-mA:
- 0.01…0.5 ppm
- 0.02…2.0 ppm
- 0.10…10.0 ppm
- CDR 1-CAN:
- 0.01…10.0 mg/l
CDR 1-CAN Specifics
- Includes 100 ml electrolyte, connecting cable (CAN M12 5-pin 0.5 m), T-distributor (M12 5-pin CAN).
Disclaimer: Extraction may miss figures, footnotes, or revisions; contractual data must match the OEM PDF revision used on the project.
Official datasheet (PDF)
Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.