Back to Prominent Sensors
ProMinent logo

Datasheet summary

DULCOTEST CLB

Engineering summary from PDF text extraction for DULCOTEST CLB. Verify every value with the OEM datasheet.

Summary

Product Overview

DULCOTEST® sensors offer reliable online measurement of free (effective) chlorine for various water applications. These amperometric sensors are available in different versions to suit a wide range of water qualities, disinfection processes, and measurement ranges.

General Features

  • Measurement Principle: Amperometric with 2 or 3 electrodes, membrane-covered (except for CLO series).
  • Output Signal: Varies by model, including 4-20 mA, 0-1 V DC, and digital (CANopen).
  • Temperature Compensation: Integrated in most models, except for CLE 3-DMT and CLB series.
  • Cross-Sensitivity: Generally low for combined chlorine (chloramines), with specific models (CLE3.1, CLE 3.1-CAN-P) designed for higher selectivity.
  • Disinfection Processes: Compatible with chlorine gas, hypochlorite, and electrolysis (except for certain models and configurations).
  • Installation: Typically in a bypass with an open sample water outlet, or inline with specific fittings.
  • Sensor Fitting: Common fittings include DGM and DLG III.
  • Electrolyte: Supplied with 100 ml of electrolyte; diaphragm caps are replaceable on relevant models.

Specific Sensor Models

CLE 3-mA Series

Standard sensors for clear water, operating with controllers featuring a 4-20 mA input.

  • Measured Variable: Free chlorine.
  • Reference Method: DPD1.
  • pH Range: 5.5 – 8.0.
  • Temperature Range: 5 – 45 °C.
  • Max. Pressure: 1.0 bar.
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: 16–24 V DC (two-wire).
  • Output Signal: 4–20 mA (uncalibrated, temperature-compensated).
  • Disinfection Process Limitations: Unsuitable for disinfectants with organic chlorine (e.g., based on cyanuric acid).
  • Resistance: Salts, acids, alkalis; not surfactants.
  • Typical Applications: Potable water, swimming pools (surfactant-free).
  • Measuring Ranges:
    • 0.01–0.5 mg/l (CLE 3-mA-0.5 ppm)
    • 0.02–2.0 mg/l (CLE 3-mA-2 ppm)
    • 0.05–5.0 mg/l (CLE 3-mA-5 ppm)
    • 0.10–10.0 mg/l (CLE 3-mA-10 ppm)
    • 0.20–20.0 mg/l (CLE 3-mA-20 ppm)
    • 0.50–50.0 mg/l (CLE 3-mA-50 ppm)
    • 1.00–100.0 mg/l (CLE 3-mA-100 ppm)

CLE 3.1-mA Series

Designed for clear water with higher selectivity towards combined chlorine, for use with 4-20 mA controllers.

  • Measured Variable: Free chlorine (HOCl), with high levels of combined chlorine.
  • pH Range: 5.5 – 8.0.
  • Temperature Range: 5 – 45 °C.
  • Max. Pressure: 1.0 bar.
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: 16–24 V DC (two-wire).
  • Output Signal: 4–20 mA (uncalibrated, temperature-compensated).
  • Selectivity: High selectivity against combined chlorine, even in excess.
  • Typical Applications: Potable water with higher combined chlorine, swimming pools.
  • Measuring Ranges:
    • 0.01–0.5 mg/l (CLE 3.1-mA-0.5 ppm)
    • 0.02–2.0 mg/l (CLE 3.1-mA-2 ppm)
    • 0.05–5.0 mg/l (CLE 3.1-mA-5 ppm)
    • 0.10–10.0 mg/l (CLE 3.1-mA-10 ppm)

CLE 3-DMT Series

Standard sensor for clear water, for operation with ProMinent transmitters type DMT.

  • Measured Variable: Free chlorine.
  • pH Range: 5.5 – 8.0.
  • Temperature Range: 5 – 45 °C.
  • Max. Pressure: 1.0 bar.
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: 3.3 V DC (5 P).
  • Output Signal: 0–1 V DC (uncalibrated, not temperature-compensated).
  • Temperature Measurement: Integrated Pt 1000 for temperature compensation in DMT transmitter.
  • Typical Applications: Potable water, swimming pools (surfactant-free).
  • Measuring Ranges:
    • 0.01–5.0 mg/l (CLE 3-DMT-5 ppm)
    • 0.10–50.0 mg/l (CLE 3-DMT-50 ppm)

CLE 3-CAN-P Series

Sensor for clear water with CAN-bus connection.

  • Measured Variable: Free chlorine.
  • pH Range: 5.5 – 8.0.
  • Temperature Range: 5 – 45 °C.
  • Max. Pressure: 1.0 bar.
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: Via CAN interface (11 - 30 V).
  • Output Signal: Digital (CANopen), uncalibrated, temperature compensated, electrically isolated.
  • Typical Applications: Potable water, swimming pool water.
  • Measuring Range:
    • 0.01–10.0 mg/l (CLE 3-CAN-P-10 ppm)

CLE 3.1-CAN-P Series

Sensor for clear water with higher selectivity to combined chlorine, for CAN-bus controllers.

  • Measured Variable: Free chlorine with high levels of combined chlorine.
  • pH Range: 5.5 – 8.0.
  • Temperature Range: 5 – 45 °C.
  • Max. Pressure: 1.0 bar.
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: Via CAN interface (11 – 30 V).
  • Output Signal: Digital (CANopen), uncalibrated, temperature compensated, galvanically isolated.
  • Typical Applications: Potable water with higher combined chlorine, swimming pools.
  • Measuring Range:
    • 0.01–10.0 mg/l (CLE 3.1-CAN-P-10 ppm)

CLO 1-mA Series

Suitable for clear water and electrolysis processes, with optional hydrodynamic cleaning for film-forming water.

  • Measured Variable: Free chlorine.
  • pH Range: 5.0 – 9.0.
  • Temperature Range: 5 – 45 °C.
  • Max. Pressure: 8.0 bar (at 25 °C).
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: 16–24 V DC (2-wire).
  • Output Signal: 4–20 mA (uncalibrated, temperature-compensated).
  • Disinfection Process: Suitable for diaphragm-free electrolysis.
  • Resistance: Surfactants, films (with hydrodynamic cleaning).
  • Measuring Principle: Amperometric, 3 electrodes, without diaphragm.
  • Measuring Ranges:
    • 0.02–2.0 mg/l (CLO 1-mA-2 ppm)
    • 0.10–10.0 mg/l (CLO 1-mA-10 ppm)

CLO 1-CAN-P Series

For clear water and electrolysis processes, with CAN-bus connection and optional hydrodynamic cleaning.

  • Measured Variable: Free chlorine.
  • pH Range: 5.0 – 9.0.
  • Temperature Range: 5 – 45 °C.
  • Max. Pressure: 8.0 bar (at 25 °C).
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: 11–30 V (via CAN interface).
  • Output Signal: Digital (CANopen), uncalibrated, temperature-compensated, galvanically isolated.
  • Disinfection Process: Suitable for diaphragm-free electrolysis.
  • Resistance: Salts, acids, lyes, surfactants, dirt films, films (with hydrodynamic cleaning).
  • Measuring Principle: Amperometric, 3 electrodes, without diaphragm.
  • Measuring Range:
    • 0.10–10.0 mg/l (CLO 1-CAN-P-10 ppm)

CLO 2-mA Series

Designed for hot water (up to 70 °C) and electrolysis processes, with optional hydrodynamic cleaning.

  • Measured Variable: Free chlorine.
  • pH Range: 5.0 – 9.0.
  • Temperature Range: 5 – 70 °C.
  • Max. Pressure: 8.0 bar (at 25 °C).
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: 16–24 V DC (2-wire).
  • Output Signal: 4–20 mA (uncalibrated, temperature-compensated).
  • Disinfection Process: Suitable for diaphragm-free electrolysis.
  • Resistance: Surfactants, films (with hydrodynamic cleaning).
  • Measuring Principle: Amperometric, 3 electrodes, without diaphragm.
  • Measuring Range:
    • 0.02–2.0 mg/l (CLO 2-mA-2 ppm)

CLB 2-µA Series

Cost-effective, simple sensor for clear water and electrolysis processes, for use with compact controllers.

  • Measured Variable: Free chlorine.
  • pH Range: 5.0 – 9.0.
  • Temperature Range: 5 – 45 °C.
  • Max. Pressure: 3.0 bar.
  • Intake Flow: 30–60 l/h.
  • Output Signal: Non-amplified primary current signal (not temperature-compensated, uncalibrated).
  • Temperature Measurement: Integrated Pt 1000.
  • Disinfection Process: Suitable for diaphragm-free electrolysis.
  • Resistance: Surfactants.
  • Measuring Principle: Amperometric, 3 electrodes, without diaphragm.
  • Measuring Range: 0.05–5.0 mg/l (up to 10 mg/l for short-term shock chlorination).

CLB 3-µA Series

Similar to CLB 2-µA but requires constant media temperature.

  • Measured Variable: Free chlorine.
  • pH Range: 5.0 – 9.0.
  • Temperature Range: 5 – 45 °C (constant temperature required).
  • Max. Pressure: 3.0 bar.
  • Intake Flow: 30–60 l/h.
  • Output Signal: Non-amplified primary current signal (not temperature-compensated, uncalibrated).
  • Temperature Measurement: None.
  • Disinfection Process: Suitable for diaphragm-free electrolysis.
  • Resistance: Surfactants.
  • Measuring Principle: Amperometric, 3 electrodes, without diaphragm.
  • Measuring Range: 0.05–5.0 mg/l (up to 10.0 mg/l for shock chlorination).

CBR 1-mA Series

For free chlorine and bromine in contaminated water, suitable for high pH up to 9.5.

  • Measured Variable: Free chlorine, free bromine, combined bromine, DBDMH.
  • Reference Method: DPD1.
  • pH Range: 5 – 9.5.
  • Temperature Range: 1 – 40 °C.
  • Max. Pressure: 1.0 bar.
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: 16–24 V DC (2-wire).
  • Output Signal: 4–20 mA (uncalibrated, temperature-compensated).
  • Resistance: Salts, acids, alkalis, surfactants, dirt films.
  • Measuring Principle: Amperometric, 2 electrodes, membrane-covered.
  • Measuring Ranges:
    • 0.01–0.5 mg/l (CBR 1-mA-0.5 ppm) - Note: Measuring range for bromine is 2.25x higher.
    • 0.02–2.0 mg/l (CBR 1-mA-2 ppm) - Note: Measuring range for bromine is 2.25x higher.
    • 0.05–5.0 mg/l (CBR 1-mA-5 ppm) - Note: Measuring range for bromine is 2.25x higher.
    • 0.10–10.0 mg/l (CBR 1-mA-10 ppm) - Note: Measuring range for bromine is 2.25x higher.
  • Typical Applications: Cooling water, process water, waste water, contaminated swimming pool water, raw water for drinking water treatment.

CLR 1-mA Series

For high concentrations of free chlorine (up to 1,000 ppm) in contaminated washing water.

  • Measured Variable: Free chlorine.
  • Reference Method: DPD1.
  • pH Range: 5.5 – 8.0.
  • Temperature Range: 5 – 45 °C.
  • Max. Pressure: 1.0 bar.
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: 16–24 V DC (2-wire).
  • Output Signal: 4–20 mA (uncalibrated, temperature-compensated).
  • Resistance: Salts, acids, alkalis, surfactants, dirt films.
  • Measuring Principle: Amperometric, 2 electrodes, membrane-covered.
  • Measuring Ranges:
    • 10.0–200 mg/l (CLR 1-mA-200 ppm)
    • 10.0–1,000 mg/l (on request)
  • Typical Applications: Salad, vegetable, and poultry washing water; contaminated process and waste water.

CBR 1-CAN-P Series

For free chlorine and bromine in contaminated water, with CAN-bus connection.

  • Measured Variable: Free chlorine, free bromine, combined bromine, DBDMH.
  • Reference Method: DPD1.
  • pH Range: 5 – 9.5.
  • Temperature Range: 1 – 40 °C.
  • Max. Pressure: 1.0 bar.
  • Intake Flow: 30–60 l/h.
  • Supply Voltage: 11–30 V DC (via CAN interface).
  • Output Signal: Digital (CANopen), uncalibrated, temperature-compensated, galvanically isolated.
  • Resistance: Dirt films, biofilms, surfactants.
  • Measuring Principle: Amperometric, 2 electrodes, membrane-covered.
  • Measuring Range:
    • 0.01–10.0 mg/l (CBR 1-CAN-P-10ppm)
  • Typical Applications: Cooling water, process water, waste water, contaminated swimming pool water.

General Operating Data

  • General Measuring Ranges: 0.01 – 200 mg/l, special measuring range up to 1,000 mg/l.
  • Applications: Potable water, swimming pool water, cooling water, industrial water, process water, rinsing water, waste water, seawater, and salt water.

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.