Back to Prominent Sensors
ProMinent logo

Datasheet summary

DULCOTEST OZE

Engineering summary from PDF text extraction for DULCOTEST OZE. Verify every value with the OEM datasheet.

Summary

Product identification

  • Product line: DULCOTEST® Sensors for Ozone
  • Manufacturer: Prominent Sensors (implied)
  • Available sensor types: OZE3 mA, OZR1

Operating limits

  • Maximum pressure: 1.0 bar
  • Temperature range: 5 … 40 °C
  • pH range: 4.0 … 11.0

Feed water

  • Intake flow: 30…60 l/h (in DGM or DLG III)

Product water

Ozone sensor OZE3 mA

  • Application: Standard sensor for measuring ozone in clear water.
  • Measured variable: Ozone (O₃)
  • Reference method: DPD4
  • Cross sensitivity: Chlorine dioxide
  • Selectivity: Ozone as against free chlorine, combined chlorine, hydrogen peroxide
  • Resistance to: Salts, acids, alkalis. Not surfactants.
  • Measuring principle: Amperometric, 2 electrodes, membrane-covered
  • Measuring range: 0.02…2.0 mg/l
  • Typical applications: Potable water and swimming pool water.

Ozone sensor OZR1 mA

  • Application: Sensor for measuring and monitoring the absence of ozone, suitable for use in contaminated water.
  • Measured variable: Ozone (O₃)
  • Reference method: DPD4
  • Cross sensitivity: Chlorine dioxide, peracetic acid, bromine, bromamine
  • Selectivity: Non-selective
  • Response time t90 (after 1 month with 0.00 ppm ozone): <210s
  • Resistance to: Salts, acids, alkalis, surfactants, dirt films
  • Measuring principle: Amperometric, 2 electrodes, membrane-covered
  • Measuring range: 0.02…2.0 mg/l
  • Typical applications: Potable water, swimming pool water, process, service or cooling water, monitoring the ozone breakdown of filters.

Electrical

  • Supply voltage: 16...24 V DC (two-wire technology/system)
  • Output signal: 4…20 mA ≈ measuring range, temperature-compensated, uncalibrated, not electrically isolated

Mechanical / hydraulic

  • Sensor fitting: DGM, DLG III
  • Installation: Bypass: open sample water outlet
  • Measuring and control equipment: D1C
  • Required for initial fitting in DLG III: Mounting kit (order no. 815079)

Disclaimer This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.