Datasheet summary
DULCOTEST CGE
Engineering summary from PDF text extraction for DULCOTEST CGE. Verify every value with the OEM datasheet.
Summary
Product identification
The DULCOTEST® CGE sensors are designed for the online measurement of total available chlorine. Two models are described:
- CGE 3-mA: For operation with controllers with a 4-20 mA input.
- CGE 3-CAN-P: For use with controllers featuring a CAN-bus connection.
Operating limits
- pH range: 5.5 … 9.5
- Temperature: 5 … 45 °C
- Max. pressure: 3.0 bar
- Intake flow: 30…60 l/h (in DGM or DLG III)
Feed water
- Disinfection process: Disinfectants with organic chlorine (e.g., based on cyanuric acid), chlorine gas, hypochlorite, electrolysis.
- Resistance to: Surfactants, cyanuric acid.
- Water quality: Similar to potable water with a pH up to 9.5.
Product water
- Measured variable: Total available chlorine (total of organically bound chlorine, e.g., bound to cyanuric acid, and free chlorine). Can also be used as a sensor for free chlorine without interference from cyanuric acid.
- Reference method: DPD1
- Selectivity: Total available chlorine and free chlorine as against combined chlorine (chloramines).
Electrical
CGE 3-mA:
- Supply voltage: 16...24 V DC (2-wire system)
- Output signal: 4-20 mA (measuring range, temperature-compensated, uncalibrated, not electrically isolated)
CGE 3-CAN-P:
- Supply voltage: Via CAN interface (11 – 30 V DC)
- Output signal: Uncalibrated, temperature-compensated, electrically isolated
Mechanical / hydraulic
- Measuring principle, technology: Amperometric, 2 electrodes, membrane-covered.
- Sensor fitting: DGM, DLG III.
- Installation: Bypass: open sample water outlet.
- Mounting kit: A mounting kit (order no. 815079) is required for initial fitting in the in-line probe housing DLG III.
Measuring ranges
- CGE 3-mA-2 ppm: 0.02…2.0 mg/l (Order no. 1047959)
- CGE 3-mA-10 ppm: 0.10…10.0 mg/l (Order no. 1047975)
- CGE 3-CAN-P-10 ppm: 0.01…10.0 mg/l (Order no. 1083211)
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.