Back to Prominent Sensors
ProMinent logo

Datasheet summary

DULCOTEST CTE

Engineering summary from PDF text extraction for DULCOTEST CTE. Verify every value with the OEM datasheet.

Summary

Product identification

DULCOTEST® Sensors for Total Chlorine, CTE 1 range.

General description

These sensors provide reliable online measurement of total chlorine, which includes free chlorine (HOCl, OCl⁻) and combined chlorine (chloramines). They are suitable for various water qualities and disinfection processes, including use at high pH values up to 9.5. The amperometric measurement offers real-time, precise readings with a short response time.

Fields of application

  • Potable water
  • Swimming pool water
  • Industrial water
  • Process water
  • Waste water

Applications include chloramination, chlorine elimination, waste water monitoring, and in swimming pools, detecting combined chlorine when used with the CLE 3.1 sensor.

Technical Data

Sensor for Total Chlorine CTE 1-mA

  • Measured variable: Total chlorine, chlorine compounds acting as oxidising agents (e.g., free chlorine, chloramines).
  • Reference method: DPD4
  • pH range: 5.5 … 9.5
  • Temperature range: 5 … 45 °C
  • Max. pressure: 3.0 bar
  • Intake flow: 30…60 l/h (in DGM or DLG III)
  • Supply voltage: 16...24 V DC (two-wire technology)
  • Output signal: 4…20 mA, temperature-compensated, uncalibrated, not electrically isolated.
  • Selectivity: Non-selective, cross-sensitive to many oxidizing agents.
  • Disinfection process compatibility: Chlorine gas, hypochlorite, electrolysis with diaphragm, monochloramine.
  • Measuring principle, technology: Amperometric, 2 electrodes, membrane-covered.

Measuring ranges

  • CTE 1-mA-0.5 ppm: 0.01…0.5 mg/l (Order no. 740686)
  • CTE 1-mA-2 ppm: 0.02…2.0 mg/l (Order no. 740685)
  • CTE 1-mA-10 ppm: 0.10…10.0 mg/l (Order no. 740684)

Sensor for Total Chlorine CTE 1-DMT

  • Measured variable: Total chlorine, chlorine compounds acting as oxidising agents (e.g., free chlorine, chloramines).
  • Reference method: DPD4
  • pH range: 5.5 … 9.5
  • Temperature range: 5 … 45 °C
  • Max. pressure: 3.0 bar
  • Intake flow: 30…60 l/h (in DGM or DLG III)
  • Supply voltage: 3.3 V DC (5 P)
  • Output signal: Uncalibrated, not temperature-compensated, not electrically isolated.
  • Selectivity: Non-selective, cross-sensitive to many oxidizing agents.
  • Disinfection process compatibility: Chlorine gas, hypochlorite, electrolysis with diaphragm, monochloramine.
  • Measuring principle, technology: Amperometric, 2 electrodes, membrane-covered.

Measuring ranges

  • CTE 1-DMT-10 ppm: 0.01…10.0 mg/l (Order no. 1007540)

Sensor for Total Chlorine CTE 1-CAN-P

  • Measured variable: Total chlorine, chlorine compounds acting as oxidising agents (e.g., free chlorine, chloramines).
  • Reference method: DPD4
  • pH range: 5.5 … 9.5
  • Temperature range: 5 … 45 °C
  • Max. pressure: 3.0 bar
  • Intake flow: 30…60 l/h (in DGMa or DLG III)
  • Supply voltage: Via CAN interface (11 - 30 V)
  • Output signal: Uncalibrated, temperature-compensated, electrically isolated.
  • Selectivity: Non-selective, cross-sensitive to many oxidizing agents.
  • Disinfection process compatibility: Chlorine gas, hypochlorite, electrolysis with diaphragm, monochloramine.
  • Measuring principle, technology: Amperometric, 2 electrodes, membrane-covered.

Measuring ranges

  • CTE 1-CAN-P-10 ppm: 0.01…10.0 mg/l (Order no. 1083210)

General Features

  • Diaphragm-covered electrodes reduce dependence on flow and water composition.
  • Integrated temperature compensation (except for CTE 1-DMT).
  • Long service life due to electrolyte-filled electrodes.

Installation

  • Installation: Bypass with open sample water outlet.
  • Sensor fitting: DGM, DLG III.
  • Mounting kit: Order no. 815079 is required for initial fitting in the DLG III probe housing.

Disclaimer

This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.

Official datasheet (PDF)

Lenntech datasheet mirror

Curated from selected public technical reference material for discovery and preliminary comparison. This summary is not a substitute for a current certified manufacturer datasheet. Verify revisions and design limits before use.