Testomat 2000 CLT
Engineering summary from PDF text extraction for Testomat 2000 CLT. Verify every value with the OEM datasheet.
Summary
Product identification
Testomat 2000® CLT is a wet-chemical online instrument for monitoring total chlorine (CLT).
Application
- Monitoring of chlorination systems for drinking water/swimming pool water
- Protection for reverse osmosis membranes
- Monitoring of biocides and conditioning agents containing chlorine
Operating limits
- Ambient temperature: 10–45 °C (50 - 104°F)
- Water temperature: 10 to 40 °C (50 to 104°F)
Feed water
- Operating pressure:
- 1 to 8 bar (15 to 116 psi)
- 0.3 to 1 bar (5 to 15 psi), after removing the valve body out of the regulator/filter housing
- Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter
Product water
- Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter
Measuring range
- Total chlorine: 0.00 - 0.99 mg/l and 1.0 - 2.5 mg/l
- Free chlorine: 0.00 - 0.99 mg/l and 1.0 - 2.5 mg/l
Electrical
- Power supply: 230–240 VAC, 115 VAC, 24 VAC (all 50–60Hz)
- Power consumption: max. 30 VA
- Protection class: I
- Degree of protection: IP 65
Mechanical / hydraulic
- Dimensions (W x H x D): 380 x 480 x 280 mm
- Weight: approx. 9.5 kg
- Menu languages: German, English, French, Italian (others upon request)
- Option: booster pump (see accessories)