4040-TS80-TSA

Engineering summary from PDF text extraction for 4040-TS80-TSA. Verify every value with the OEM datasheet.

Summary

Product identification

The TRISEP® TS80 series consists of High Rejection NF (nanofiltration) Elements. These nanofiltration membranes offer high solute rejection for both salts and uncharged organic solutes, operating at lower pressures compared to reverse osmosis membranes. The TS80 is considered a "softening" membrane for many water purification applications, typically operating at about 100 psi. The membrane material is Polyamide.

Membrane Characteristics

  • Typical NaCl rejection: 80%
  • Stabilized MgSO₄ rejection: 99.2%
  • Minimum MgSO₄ rejection: 98.5%

Operating limits

  • Maximum Operating Pressure: 41 bar (600 psi)
  • Maximum Operating Temperature: 45°C (113°F)
  • Cleaning pH Range: 1.0 – 12.0
  • Chlorine Tolerance: < 0.1 ppm
  • Maximum Pressure Drop: 1 bar (15 psi) per element; 4 bar (60 psi) per housing
  • Maximum SDI₁₅: 5.0
  • Maximum Turbidity: 1 NTU

Design Information

Test conditions for permeate flow rates: 2,000 ppm MgSO₄, 7.6 bar (110 psi), 25°C (77°F), 15% recovery, pH 8.0, 30 minutes operation. Flow rates will be no more than 15% below the values shown. All models on this sheet have fiberglass outer wrap. The TRISEP® 8040-TS80-TXFA has parallel feed spacers; all other models have diamond-shaped feed spacers. All models include anti-telescoping devices (ATDs) attached to the ends of the element and one brine seal. All 4040 and 8040 models include one interconnector.

ModelPart NumberPermeate Flow m³/day (GPD)Membrane Area m² (ft²)Feed Spacer Thickness (mil)
TRISEP® PLT 2540-TS80-311630002422.3 (600)2.4 (26)31
TRISEP® PLT 4040-TS80-311610204438.0 (2,100)7.9 (85)31
TRISEP® 4040-TS80-TSA1630004428.0 (2,100)8.2 (88)31
TRISEP® 8040-TS80-TSA16300084236.0 (9,500)33.9 (365)34
TRISEP® 8040-TS80-TSFA16804084236.0 (9,500)33.9 (365)34
TRISEP® 8040-TS80-TXFA19102084026.5 (7,000)25.3 (275)47
TRISEP® 8040-TS80-UWA16401184239.7 (10,500)37.2 (400)28
TRISEP® 8040-TS80-UWFA16401284239.7 (10,500)37.2 (400)31

Mechanical / Hydraulic

For Female elements, Dimension “C” is the Inner Diameter. For Male elements, “C” is the Outer Diameter. Male elements have a protruding permeate tube, indicated as “D” in the diagram (not included in extract). Dimension “D” is 25.4 mm (1.0 in).

ModelElement Weight kg (lb)Dim. A mm (inches)Dim. B mm (inches)Dim. C mm (inches)Permeate Tube
TRISEP® PLT 2540-TS80-313 (7)1,016 (40.0)64 (2.5)19.1 (0.75)Male
TRISEP® PLT 4040-TS80-314 (9)1,016 (40.0)99 (3.9)19.1 (0.75)Male
TRISEP® 4040-TS80-TSA4 (9)1,016 (40.0)99 (3.9)19.1 (0.75)Female
TRISEP® 8040-TS80-TSA16 (36)1,016 (40.0)201 (7.9)38.1 (1.50)Female
TRISEP® 8040-TS80-TSFA16 (36)1,016 (40.0)201 (7.9)28.6 (1.125)Female
TRISEP® 8040-TS80-TXFA16 (36)1,016 (40.0)201 (7.9)28.6 (1.125)Female
TRISEP® 8040-TS80-UWA16 (36)1,016 (40.0)201 (7.9)38.1 (1.50)Female
TRISEP® 8040-TS80-UWFA16 (36)1,016 (40.0)201 (7.9)28.6 (1.125)Female

Important Information

  • Start-up: Flush elements for 30 minutes at low pressure and discard permeate during the flush prior to operation.
  • Cleaning: Membrane elements require periodic cleaning to maintain proper operation and prevent damage. Refer to the Membrane Cleaning Guide – Water Application Elements (TSG-C-001) for detailed instructions.
  • Storage: Elements must be stored appropriately to ensure proper operation and prevent damage. Refer to Element Storage Guides (TSG-O-009 & TSG-O-010).

Disclaimer: This summary is based on extracted text and may not include all figures, footnotes, or the latest revisions. Contractual data must match the OEM PDF revision used on the project.