Back to Wallace & Tiernan Analyzers - Controllers
Wallace & Tiernan logo

Datasheet summary

Membran Sensors FC2, TC3, CD7, OZ7

Engineering summary from PDF text extraction for Membran Sensors FC2, TC3, CD7, OZ7. Verify every value with the OEM datasheet.

Summary

Product Identification

Wallace & Tiernan® membrane sensor measuring modules are available for:

  • Free Chlorine (FC2)
  • Total Chlorine (TC3)
  • Chlorine Dioxide (CD7)
  • Ozone (OZ7)

These are compatible with SFC and MFC electronics and various flow cell modules.

Benefits of Membrane Sensors

  • Safe, long-term stable measurement system
  • Integral temperature compensation
  • No reagents required

Benefits of FC2 for Free Chlorine

  • Continuous, on-line measurement of free chlorine even at high pH (up to pH 10) and low conductivity (from 10 μS/cm).
  • Ideal for samples with varying pH due to internal buffering.

Typical Applications

  • Measurement and control in industrial and municipal water treatment.
  • Process measurement, final inspections, and neutralization plants.
  • Monitoring of potable water, well water, pool water, and boiler feed water.

Technical Data

Sensor

FeatureFC2/TC3CD7/OZ7
Measuring SystemMembrane-covered 3-electrode potentiostaticMembrane-covered 2-electrode potentiostatic
Measured ParameterFree Chlorine / Total ChlorineChlorine Dioxide / Ozone
ElectrolyteDiluted potassium halide solutionDiluted potassium halide solution
Reference ElectrodeSilver/Silver halide/potassium halide solutionSilver/Silver halide/potassium halide solution
ApplicationClear water (potable, pool, brine)All kinds of water (wastewater, process, salt/brine, water with tensides)
Dimensionsø 25 mm, length 175 mmø 25 mm, length 175 mm
Models (Pressure)FC2: up to 3 bar; TC3: up to 0.5 barUp to 1 bar
Measuring Range0.05 to 20 ppm (TC3: includes chloramines)CD7: 0.05 to 20 ppm / OZ7: 0.02 to 10 ppm
Response Time (T90)FC2: 60 to 300 s; TC3: up to 5 minCD7: approx. 20 s; OZ7: > 50 s
Temperature Range+5 to +45 °C+5 to +45 °C
Conductivity10 μS/cm to 50 mS/cm> 1 μS/cm to 40 mS/cm

Flow Module

DEPOLOX® 5 Flow-Through Adapter

  • Integrated, open sensor for oxidation and disinfection chemicals.
  • Compatible with additional measurements of MFC/SFC series.
  • Sample water flow: Controlled to 33 l/h (0.15 US gpm) with max. 4 bar (58 psi) inlet pressure*.
  • Integrated multi-sensor with flow monitor and compatible with temperature sensor.
  • Max. sample water temperature: +50 °C (122 °F).

VariaSens Flow-Through Adapter

  • For use in combination with membrane sensors and additional measurements of the MFC/SFC series.
  • Sample water flow: Controlled to 33 l/h (0.15 US gpm) with max. 4 bar (58 psi) inlet pressure*.
  • Integrated multi-sensor with flow monitor and compatible with temperature sensor.
  • Max. sample water temperature: +50 °C (122 °F).
  • CD7/OZ7: Pressurized applicable with up to 1 bar back pressure in VariaSens module.

*Note: Sample water pressures of up to 40 bar (580 psi) can be adapted with special equipment.

Electronic

  • Sensor card for MFC and SFC.
  • Measurement Input: For membrane sensor, pre-calibrated.
  • Input Data: -7 to 1000 μA.
  • Accuracy: 0.5 % FS.
  • Temperature Drift: < 0.2 %/10 K.

Cable

  • Supplied with a 0.9 m (3.0 ft.) screened coaxial cable.
  • Extension cables up to 50 m (164 ft.) are available as optional accessories.
  • Utilizing "plug & play" technology allows automatic sensor recognition.
  • Analog output (0/4 to 20 mA) and user-configurable alarm contacts are available.

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.