Datasheet summary
Membran Sensors FC2, TC3, CD7, OZ7
Engineering summary from PDF text extraction for Membran Sensors FC2, TC3, CD7, OZ7. Verify every value with the OEM datasheet.
Summary
Product Identification
Wallace & Tiernan® membrane sensor measuring modules are available for:
- Free Chlorine (FC2)
- Total Chlorine (TC3)
- Chlorine Dioxide (CD7)
- Ozone (OZ7)
These are compatible with SFC and MFC electronics and various flow cell modules.
Benefits of Membrane Sensors
- Safe, long-term stable measurement system
- Integral temperature compensation
- No reagents required
Benefits of FC2 for Free Chlorine
- Continuous, on-line measurement of free chlorine even at high pH (up to pH 10) and low conductivity (from 10 μS/cm).
- Ideal for samples with varying pH due to internal buffering.
Typical Applications
- Measurement and control in industrial and municipal water treatment.
- Process measurement, final inspections, and neutralization plants.
- Monitoring of potable water, well water, pool water, and boiler feed water.
Technical Data
Sensor
| Feature | FC2/TC3 | CD7/OZ7 |
|---|---|---|
| Measuring System | Membrane-covered 3-electrode potentiostatic | Membrane-covered 2-electrode potentiostatic |
| Measured Parameter | Free Chlorine / Total Chlorine | Chlorine Dioxide / Ozone |
| Electrolyte | Diluted potassium halide solution | Diluted potassium halide solution |
| Reference Electrode | Silver/Silver halide/potassium halide solution | Silver/Silver halide/potassium halide solution |
| Application | Clear water (potable, pool, brine) | All kinds of water (wastewater, process, salt/brine, water with tensides) |
| Dimensions | ø 25 mm, length 175 mm | ø 25 mm, length 175 mm |
| Models (Pressure) | FC2: up to 3 bar; TC3: up to 0.5 bar | Up to 1 bar |
| Measuring Range | 0.05 to 20 ppm (TC3: includes chloramines) | CD7: 0.05 to 20 ppm / OZ7: 0.02 to 10 ppm |
| Response Time (T90) | FC2: 60 to 300 s; TC3: up to 5 min | CD7: approx. 20 s; OZ7: > 50 s |
| Temperature Range | +5 to +45 °C | +5 to +45 °C |
| Conductivity | 10 μS/cm to 50 mS/cm | > 1 μS/cm to 40 mS/cm |
Flow Module
DEPOLOX® 5 Flow-Through Adapter
- Integrated, open sensor for oxidation and disinfection chemicals.
- Compatible with additional measurements of MFC/SFC series.
- Sample water flow: Controlled to 33 l/h (0.15 US gpm) with max. 4 bar (58 psi) inlet pressure*.
- Integrated multi-sensor with flow monitor and compatible with temperature sensor.
- Max. sample water temperature: +50 °C (122 °F).
VariaSens Flow-Through Adapter
- For use in combination with membrane sensors and additional measurements of the MFC/SFC series.
- Sample water flow: Controlled to 33 l/h (0.15 US gpm) with max. 4 bar (58 psi) inlet pressure*.
- Integrated multi-sensor with flow monitor and compatible with temperature sensor.
- Max. sample water temperature: +50 °C (122 °F).
- CD7/OZ7: Pressurized applicable with up to 1 bar back pressure in VariaSens module.
*Note: Sample water pressures of up to 40 bar (580 psi) can be adapted with special equipment.
Electronic
- Sensor card for MFC and SFC.
- Measurement Input: For membrane sensor, pre-calibrated.
- Input Data: -7 to 1000 μA.
- Accuracy: 0.5 % FS.
- Temperature Drift: < 0.2 %/10 K.
Cable
- Supplied with a 0.9 m (3.0 ft.) screened coaxial cable.
- Extension cables up to 50 m (164 ft.) are available as optional accessories.
- Utilizing "plug & play" technology allows automatic sensor recognition.
- Analog output (0/4 to 20 mA) and user-configurable alarm contacts are available.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.