Datasheet summary
Testomat 2000 CN
Engineering summary from PDF text extraction for Testomat 2000 CN. Verify every value with the OEM datasheet.
Summary
Product identification
The Testomat 2000® CN is an online analysis instrument for water hardness intended for the Chinese market. It features a Chinese menu and is identical to the Testomat 2000®.
Applications
- Water treatment plants
- Water blending plants
- Drinking water plants
- Water softening plants
- Decarbonization plants
- Desalination plants
- Boiler houses
- Cooling towers
Measuring range
- Water hardness: 0.05–25 °dH
- Carbonate hardness: 0.5–20 °dH
- p-value: 1–15 mmol/l
- minus m-value: 0.05–0.5 mmol/l (depending on the choice of indicator)
Technical Data
Electrical
- Power supply: 230–240 VAC, 115 VAC, 24 VAC (all 50–60 Hz)
- Power consumption: max. 30 VA
- Protection class: I
Mechanical / hydraulic
- Degree of protection: IP 65
- Dimensions (W x H x D): 380 x 480 x 280 mm
- Weight: approx. 9.5 kg
Operating limits
- Ambient temperature: 10–45 °C (50 - 104°F)
Feed water
- Operating pressure:
- 1 to 8 bar (15 to 116 psi) or
- 0.3 to 1 bar (5 to 15 psi), after removing the valve body from the regulator/filter housing
- Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter
- Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter
- Water temperature: 10 to 40 °C (50 to 104°F)
Additional features
- Menu languages: Mandarin, English
- Optional booster pump (see accessories)
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.