Datasheet summary
Testomat 2000 Polymer
Engineering summary from PDF text extraction for Testomat 2000 Polymer. Verify every value with the OEM datasheet.
Summary
Product identification
The Testomat 2000® Polymer is a process photometer and wet-chemical online monitor for polyacrylate content.
Application
- Monitoring of conditioning agents in cooling and heating circuits.
Measuring range
- 0 to 50 mg/l (ppm) with a factor set to 1.00.
- Customer-specific ranges, e.g., 0.0-50.0 mg/l.
Customization is necessary due to the range of polyacrylates measurable.
Technical Data
- Power supply: 230–240 VAC, 115 VAC, 24 VAC (50–60 Hz)
- Power consumption: max. 30 VA
- Protection class: I
- Degree of protection: IP 65
- Ambient temperature: 10–45 °C (50 - 104°F)
- Dimensions (W x H x D): 380 x 480 x 280 mm
- Weight: approx. 9.5 kg
- Menu languages: German, English (others upon request)
Operating limits
Mains water supply
- Operating pressure: 1 to 8 bar (15 to 116 psi) or 0.3 to 1 bar (5 to 15 psi) after removing the valve body from the regulator/filter housing.
- Water temperature: 10 to 40 °C (50 to 104°F)
Connections
- Water inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter.
- Water outlet: Hose with 6 mm external diameter / 4 mm internal diameter.
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.