Datasheet summary
Testomat 2000 self clean
Engineering summary from PDF text extraction for Testomat 2000 self clean. Verify every value with the OEM datasheet.
Summary
Product identification
Testomat 2000® self clean
Application
- Use for difficult water, e.g. calcium, biofilms, various other deposits
- Extending service life
- Reducing contamination in the measuring chamber
Measuring range
- 0.05–25 °dH Water hardness
- 0.5–20 °dH Carbonate hardness
- 1–15 mmol/l p-value
- 0.05–0.5 mmol/l minus m-value (depending on the choice of the indicator)
Technical Data
- Power supply: 230–240 VAC, 115 VAC, 24 VAC (all 50–60 Hz)
- Power consumption: max. 30 VA
- Protection class: I
- Degree of protection: IP 65
- Ambient temperature: 10–45 °C (50 - 104°F)
- Option: booster pump (see accessories)
- Dimensions (W x H x D): 380 x 480 x 280 mm
- Weight: approx. 9.5 kg
- Menu languages: German, English (others upon request)
Mains water supply
- Operating pressure:
- 1 to 8 bar (15 to 116 psi) or
- 0.3 to 1 bar (5 to 15 psi), after removing of the valve body out of the regulator/filter housing
- Water inlet: Opaque hose with 6 mm external diameter/4 mm internal diameter
- Water outlet: Hose with 6 mm external diameter/4 mm internal diameter
- Water temperature: 10 to 40 °C (50 to 104°F)