Summary

Product identification

Testomat 2000® SO3

This is a wet-chemical online instrument for monitoring the content of Sulphite (SO₃²⁻) in the 0 to 20 mg/l range using the colorimetric method.

Application

  • Monitoring of boiler feed water in steam boiler systems (sulfite for oxygen binding).

Measuring range

  • Sulfite: 0.0 - 9.9 mg/l
  • Sulfite: 10 - 20 mg/l

Operating limits

  • Ambient temperature: 10–45 °C

Feed water

  • Operating pressure: 1 to 8 bar (15 to 116 psi) or 0.3 to 1 bar (5 to 15 psi), after removing of the valve body out of the regulator/filter housing
  • Water inlet: Opaque hose with 6 mm outer diameter / 4 mm inner diameter
  • Water outlet: Hose with 6 mm outer diameter / 4 mm inner diameter
  • Water temperature: 10 to 40 °C

Electrical

  • Supply voltage: 230–240 VAC, 115 VAC, 24 VAC (all 50–60 Hz)
  • Power consumption: max. 30 VA
  • Protection class: I
  • Protection type: IP 65

Mechanical / hydraulic

  • Dimensions (W x H x D): 380 x 480 x 280 mm
  • Weight: approx. 9.5 kg
  • Menu languages: German, English (further on request)
  • Optional: Pressure booster pump (see Accessories)

Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.