Summary
Product Identification
- Product Name: Testomat® 808
- Type: Limit value monitoring device for water hardness
Application
- Reverse osmosis plants
- Soft water for commercial purposes
- Pure water production plants
- Galvanization
Measuring Range
- Water hardness: 0.02 – 3 °dH (0.4….54 ppm CaCO₃)
Technical Data
Electrical
- Power Supply: Depending on product configuration: 24V / 115V / 230V, 50 – 60 Hz
- Instrument Protection: 230 – 240 V: T0,1 A; 115 V: T0,2 A; 24 V: T0,8 A
- Mains Protection for Consumers: max. 4 A (n, l)
- Power Consumption: max. 16 VA (without external load)
- Protection Class: I
- Current Interface: Output of defined values (5, 8, 11, 14, 17, 20 mA) for displaying status and error messages, max. load 500 Ohms
- Contact Load Relay: 230V / 4A AC ohm resistive load
Mechanical / Hydraulic
- Dimensions: B x H x T = 364 x 314 x 138 mm
- Weight: 4350 g
- Degree of Protection: IP 54
Operating Limits
- Ambient Temperature: 10 – 40 °C; 50 - 104°F
Feed Water
- Operating Pressure: Depending on product configuration:
- 0.3 – 1 bar / 0.3 x 10⁵ to 1 x 10⁵ Pa
- 1 - 4 bar / 1 x 10⁵ to 4 x 10⁵ Pa
- (A pressure reducer (special accessory) should be used from 4 to 8 bar range)
- Water Inlet: Opaque hose with 6 mm external diameter / 4 mm internal diameter
- Water Outlet: Hose with 6 mm external diameter / 4 mm internal diameter
- Water Temperature: 10 to 40 °C; 50 to 104°F
Conformity
- EN 61000-6-2
- EN 61000-6-4
- EN 61010-1