Testomat 2000 FE

Engineering summary from PDF text extraction for Testomat 2000 FE. Verify every value with the OEM datasheet.

Summary

Product identification

The Testomat 2000® Fe is a process photometer designed for monitoring dissolved iron content (II)+(III). It operates using a colorimetric analysis principle.

Intended use

This instrument is intended for online monitoring of dissolved iron content in water, with applications in monitoring de-ironing plants or process/drinking water flows. It is designed for the measurement range of 0 to 1.0 mg/l.

Performance specifications

  • Measurement Range:
    • Range I: 0 - 0.65 mg/l, with a resolution of 0.01 mg/l
    • Range II: 0.7 – 1.0 mg/l, with a resolution of 0.1 mg/l
  • Analysis Principle: Colorimetric analysis with the addition of two reagents.
  • Reaction Time: Approximately 7 minutes (absolute measuring time, excluding flushing).
  • Analysis Initiation Options:
    • Automatic interval operation (interval pause adjustable from 0-99 minutes).
    • External control.
    • Dynamic (exhaustion-dependent interval operation).
    • Quantity dependent via impulse water meter.
  • Limit Values: Two independent limit values with adjustable hysteresis (1, 2, or 3 bad analyses) and switch functions.
  • Measuring Points: Supports monitoring of two measuring points via external solenoid valves.
  • Reagent Storage: Utilizes 500 ml reagent bottles.
  • Optional Interfaces:
    • Interface card for 0/4-20 mA or 0/2-10 V output.
    • Interface card for RS232 (protocol printer).

Operating Limits

  • Operating Pressure: 0.3 to 8 bar (factory set for 1 to 8 bar; can be adjusted for 0.3 to 1 bar).
  • Water Temperature: 10°C to 40°C. For temperatures above 40°C, a KCN type cooler is recommended.
  • pH Range: 4 – 10.5.

Feed Water

  • Inlet Pressure: Must be between 0.3 bar and 8 bar.
  • Inlet Hose: Standard connection for opaque plastic hoses 6/4 x 1 (external diameter 6 mm/internal diameter 4 mm, wall thickness 1 mm). Maximum length of 5 meters.
  • Water Meter: Used for quantity-dependent analysis start. The water meter constant can be programmed (factory set to 100 litres/impulse).

Product Water

  • Outlet Hose: Requires an outlet hose with an internal diameter of 12 mm. The outlet should be laid without backwater or siphoning effects.

Electrical

  • Power Supply: Mains voltage specified on the rating plate (e.g., 115/230 VAC or 24 V). Requires an interference-free and stable power supply.
  • Relay Outputs: Maximum load capacity must not be exceeded. Outputs are volt-free, rated for max. 240 VAC, 4 A.
    • Flushing (external flush valve)
    • LV1 and LV2 limit value outputs (normally closed, normally open, common)
    • Measuring point 1 or 2 (switch-over)
    • AUX (programmable function output)
    • Alarm (fault message output)
    • Maintenance (maintenance message output)
  • Signal Inputs:
    • Start (external analysis start)
    • Stop (external analysis stop)
    • IN1 (Universal input 1)
    • IN2 (Universal input 2 - water meter)
  • Analog/Serial Output: 0/4 - 20 mA galvanically separated or serial interface RS232.

Mechanical / Hydraulic

  • Dimensions: Not specified in the extract. Mounting is typically vertical and vibration-free.
  • Connections: Standard plug connector for water inlet hose; requires 12 mm internal diameter hose for outlet.
  • Components: Includes dosing pumps, water connections, flow controller, measuring chamber.

Reagents

  • Requires Heyl Testomat 2000® reagents FE2005A and FE2005B for trouble-free operation.
  • Caution: Observe correct allocation of reagents A (left) and B (right) to the dosing pumps.