Testomat 2000

Engineering summary from PDF text extraction for Testomat 2000. Verify every value with the OEM datasheet.

Summary

Product identification

The Testomat 2000® is an online analysis unit for determining and monitoring water hardness, carbonate hardness, p-value, or minus m-value.

Performance specifications

  • Menu-driven operation and programming via a plain text display.
  • Determines residual hardness, total hardness, carbonate hardness, minus m-value, and p-value based on indicator selection.
  • Freely selectable hardness units: °dH, °f, ppm CaCO₃, mmol/l.
  • High measuring accuracy due to a precise piston-dosing pump.
  • Analysis initiation methods:
    • Automatic interval operation (Interval pause adjustable from 0 to 99 minutes)
    • External control
    • Dynamic start (Exhaustion dependent)
    • Volume controlled (water meter)
  • Two independent limit values with hysteresis (1, 2 or 3 bad analyses) and adjustable switch functions.
  • Monitoring of two measuring points (change-over via external solenoid valves).
  • Internal error documentation.
  • Programmable service address.
  • Programmable maintenance interval for maintenance requests.
  • Extended operating periods with a 500 ml indicator storage bottle.
  • Optional interfaces:
    • 0/4-20 mA or 0/2-10 V
    • RS 232 for a log printer
    • SD Card data logger

Indicators and Measuring Ranges

Parameter/Indicator typeUnitRange (example for TH 2005)Resolution (example for TH 2005)
Water hardness°dH0.05 – 0.500.01
°f0.09 – 0.890.02
ppm CaCO₃0.89 – 8.930.2
mmol/l0.01 – 0.090.01
Carbonate hardness°dH0.5 – 5.00.5
°f0.90 – 8.960.9
ppm CaCO₃8.9 - 89.58.9
mmol/l0.18 – 1.790.18
Minus m-valuemmol/l0.05 – 0.500.01
p-valuemmol/l1 - 151
  • Values outside the measuring range are indicated by "<" (below) or ">" (above) the specified limits.
  • The device does not monitor dosage when TC indicators are used for carbonate hardness.

Application instructions

  • Operates within a pH range of 4 – 10.5 when determining total hardness.
  • Heavy metal ions (iron > 0.5 mg/l, copper > 0.1 mg/l, aluminum > 0.1 mg/l) may influence color reactions.
  • Incorrect evaluations may occur if the measuring water contains more than 20 mg/l CO₂.
  • Visual inspection at regular intervals is recommended for operational reliability.

Installation

Operating Limits

  • Water Pressure: 0.3 bar to 8 bar (factory setting is 1 to 8 bar). For 0.3 to 1 bar operation, a flow controller valve body must be removed.
  • Water Temperature: 10 °C to 40 °C. A cooler is recommended for temperatures above 40 °C.

Water Inlet

  • Uses a standard plug connector for plastic hoses 6/4 x 1.
  • Recommended maximum hose length: 5 m.
  • The branch line connection should be laid vertically upwards to prevent dirt ingress.
  • An opaque plastic hose is recommended.
  • For operation between 0.3 to 1 bar or with a booster pump, the valve body from the controller and filter housing must be removed. The pump should have a feeding capacity of 25 to 35 litres/hour.

Water Outlet

  • Requires an outlet hose with an internal diameter of 12 mm.
  • The hose must be laid without backwater development or siphoning effects.

Electrical Connections

  • Mains voltage: 24 V / 115 V / 230 V.
  • Connections for mains input, neutral conductor, and phase.
  • Switched mains output for consumers, max. 4 A.
  • Protective earth conductor connection (PE).
  • Protect against interference voltages using a mains filter and separate connecting cables from power cables.

Signal Inputs/Outputs

TerminalTypeFunctionComment
1-2OUTExternal flush valveFloating output, max. 240 VAC, 4 A
3-5OUTLimit value output 1Floating output, max. 240 VAC, 4 A
6-8OUTLimit value output 2Floating output, max. 240 VAC, 4 A
9-11OUTMeasuring point 1/2 switchFloating output, max. 240 VAC, 4 A
12-13OUTUniversal output (AUX)Floating output, max. 240 VAC, 4 A
14-16OUTFault message outputFloating output, max. 240 VAC, 4 A
17-19OUTMaintenance messageFloating output, max. 240 VAC, 4 A
20-21INExternal analysis startIsolated normally open contact required
22-23INExternal analysis stopIsolated normally closed/open contact required
24-25INUniversal input 1Isolated normally closed/open contact required
26-27INUniversal input 2 (water meter)Isolated normally open contact required
28-29Interface (optional)RS232, 0/4-20mA, 0/2-10V

Commissioning

  • Use only Heyl Testomat 2000® indicators.
  • Ensure the intake hose and transport hose are filled with indicator up to the measuring chamber.
  • The measuring chamber should fill within 2 to 6 seconds.

Functions of the operating and display elements

  • Mains Switch: To turn the instrument on or off.
  • Instrument Fuse: Protects the instrument and outputs against overloads.
  • Display: Shows operating statuses, measured values, limit values, error messages, and maintenance requests.
    • Limit value displays (red/green) indicate if limit values are reached/exceeded or fallen below.
    • Text display (4 lines) shows current results, statuses, and programming data.
    • Alarm indicator (red) for malfunctions/errors.
    • Maintenance indicator (yellow) for maintenance requests.
    • Status displays for active instrument components.
  • Operating Keys:
    • Manual key: Starts an analysis manually.
    • STANDBY key: Puts the instrument into standby mode.
    • Horn key: Acknowledges error and warning messages.
    • M key: Opens the programming menu.
    • i key: Calls up instrument information and settings.
  • Programming Keys (Cursor Block): Used for navigating menus, selecting functions, and entering data.

Password protection and basic program

A four-digit password, which is the current time in reverse order, is required to access the basic program for data entry and settings.

Entering basic program data

  • Indicator Type and Bottle Size: Select the indicator type and bottle size (e.g., 100ml or 500ml).
  • Mode of Operation: Choose between:
    • Time Controlled (set interval pause and flush times)
    • Volume Interval (set flow rate in litres)
    • Volume control / Time priority
    • Dynamics (filter capacity dependent control, set capacity, start/end intervals, and reset condition)
    • External Start
  • Display Unit: Select the desired unit for displayed values (°dH, °f, ppm CaCO₃, mmol/l).
  • Internal Flushing: Recommended for longer downtimes or analysis intervals to ensure representative samples. The flushing time affects the actual analysis interval.

Disclaimer

This extraction is based on provided text and may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.