Summary
Product identification
Titromat® M1 is used for automatically determining and monitoring carbonate hardness in low hardness measuring ranges.
Applications
- Corrosion monitoring in boiler feed water.
- Residual alkalinity after decarbonization (e.g., breweries).
Operating limits
Measuring range
- 0.05 - 1.00 °dH (resolution 0.025 °dH)
- 0.09 - 1.80 °f (resolution 0.045 °f)
Feed water
Operating pressure
- 1 to 8 bar (15 to 116 psi)
- 0.3 to 1 bar (5 to 15 psi) (after removing the valve body from the regulator/filter housing)
Water inlet
- Opaque hose with 6 mm external diameter / 4 mm internal diameter.
Water outlet
- Hose with 6 mm external diameter / 4 mm internal diameter.
Water temperature
- 10 to 40 °C (50 to 104°F).
Electrical
- Power supply: 230–240 VAC, 115 VAC, or 24 VAC, all 50–60 Hz.
- Power consumption: Max. 30 VA.
- Protection class: I.
Mechanical / hydraulic
- Degree of protection: IP 65.
- Ambient temperature: 10–45 °C (50 - 104°F).
- Dimensions (W x H x D): 380 x 480 x 280 mm.
- Weight: Approx. 9.5 kg.
- Menu languages: German, English, French (others upon request).
Disclaimer: This extraction may miss figures, footnotes, or revisions. Contractual data must match the OEM PDF revision used on the project.